Comparative Evaluation: e|1vet

Jul 21, 2026
What the Data Says
about the

Comparative evaluation of a new point-of-care blood electrolyte analyzer in the veterinary sector

Jennifer Lothion-Roy[a,b], Germano Ferrari[c], Julie Glenister[d], Florence Hillen[a],
Mark Dunning[a,e], Kate White[a]

[a] School of Veterinary Medicine and Science, University of Nottingham, Sutton Bonington, United Kingdom;
[b] Biodiversity Institute, University of Nottingham, Nottingham, United Kingdom; [c] EXIAS Medical GmbH, Graz, Austria;
[d] Phoenix Vets, Loughborough, United Kingdom; [e] Willows Veterinary Centre and Referral Service, Solihull, United Kingdom

Electrolyte monitoring plays a central role in companion animal medicine, supporting the diagnosis and management of metabolic, gastrointestinal, renal, and endocrine disorders. In the veterinary emergency and critical care (ECC) setting, having a point-of-care (POC) blood electrolyte analyzer that is both simple to use and fast to deliver results can make a measurable difference to patient outcomes.

A new study from the University of Nottingham, conducted in collaboration with Phoenix Vets (Loughborough, UK) and presented at EVECC 2026, evaluated exactly that: how the EXIAS e|1vet performs against two established veterinary electrolyte analyzers – the Siemens epoc Blood Analysis System and the IDEXX Catalyst One.

More about the product

Across 69 patients (45 dogs, 24 cats), the e|1vet achieved validated agreement on all six parameters measured against the epoc system – sodium, potassium, chloride, ionized calcium, pH, and hematocrit – and matched the Catalyst One closely on potassium and chloride. The one exception, sodium versus the Catalyst One, is worth noting in context: the two predicate devices only agreed with each other on sodium at 88.1%, so the discrepancy appears to sit between the established devices rather than pointing to a limitation of the e|1vet.

Just as notable were the practical advantages: the e|1vet required only 20 microliters of blood – a fraction of what the predicate devices need – and delivered results in 25 seconds, far faster than either comparator.

The authors conclude that the e|1vet is a highly accurate blood electrolyte analyzer, offering comparable to superior accuracy while requiring less blood and delivering faster results – making it a strong option for electrolyte monitoring in small animal veterinary practice.

The full study, including methodology and complete data, is available in the poster below.

Reference: Lothion-Roy J, Ferrari G, Glenister J, Hillen F, Dunning M, White K. Comparative evaluation of a new point-of-care blood electrolyte analyzer in the veterinary sector. Poster presented at: EVECC Congress; 2026 Jun; Edinburgh, UK.

Ethical approval for this study was granted by the University of Nottingham Ethics Committee (no. 4534 191125). EXIAS Medical GmbH provided the analyzer on loan for the study; sample collection and processing were carried out by the team at Phoenix Vets.